    Intended Use

    The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a qualitative in vitro test for the detection of C. trachomatis plasmid DNA in urine from males and females, in endocervical swab specimens, and in male urethral swab specimens as evidence of symptomatic or asymptomatic infection with C. trachomatis. C. trachomatis DNA is detected by Polymerase Chain Reaction (PCR) amplification of target DNA and by hybridization capture of amplified target using the COBAS AMPLICOR Analyzer.

    Device Description

    The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a multiplex in vitro diagnostic test performed on the COBAS AMPLICOR Analyzer. The COBAS AMPLICOR Analyzer automates the annolification, the nucleic acid hybridization, and the colorimetric detection procedures of the Test. The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis also has an Internal Control that identifies. specimens that contain substances inhibitory to PCR.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the COBAS AMPLICOR™ CT/NG Test for Chlamydia trachomatis, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined "acceptance criteria" in terms of specific sensitivity and specificity thresholds that needed to be met. Instead, it presents the device's performance in comparison to existing gold standards (culture and DFA). The reported performance across various patient groups and specimen types allows implicit evaluation against generally accepted standards for diagnostic tests.

    MetricAcceptance Criteria (Implicit/General)Reported Device Performance (Range)
    Analytical Sensitivity (Limit of Detection)Sufficient to detect C. trachomatis1 Inclusion Forming Unit (1 IFU) per test for all 15 Chlamydia serovars.
    Analytical SpecificityNo cross-reactivity with common urogenital flora/pathogensNegative results for all 152 bacterial, 6 fungal, 1 protozoan, and 11 viral isolates at ≥ 10^4 copies of genomic DNA per test.
    Precision (Reproducibility)Qualitatively correct results, low variability in absorbance100% qualitatively correct results across all specimen types, concentrations (0, 1.25, 3.75, 6.25 IFU/test for CTM; 0, 1, 3, 5 IFU/test for urine), and sites. Mean, Min, Max A660 values reported show consistency.
    Clinical Sensitivity (Females - individual specimen)High clinical detection rateCTM: 94.9% (93.6-96.2%) Asymptomatic; 95.9% (94.6-97.2%) Symptomatic. Urine: 90.9% (84.5-97.3%) Asymptomatic; 90.3% (84.3-96.3%) Symptomatic.
    Clinical Specificity (Females - individual specimen)High ability to correctly identify negative casesCTM: 98.3% (97.6-99.1%) Asymptomatic; 98.0% (97.1-98.8%) Symptomatic. Urine: 98.3% (97.4-99.1%) Asymptomatic; 96.9% (95.8-97.9%) Symptomatic.
    Clinical Sensitivity (Males - individual specimen)High clinical detection rateCTM: 98.7% (97.1-100%) Asymptomatic; 96.8% (94.3-99.3%) Symptomatic. Urine: 89.9% (83.2-96.5%) Asymptomatic; 88.3% (83.8-92.8%) Symptomatic.
    Clinical Specificity (Males - individual specimen)High ability to correctly identify negative casesCTM: 97.7% (96.6-98.9%) Asymptomatic; 94.6% (93.2-96.0%) Symptomatic. Urine: 96.3% (94.8-97.7%) Asymptomatic; 92.2% (90.5-93.8%) Symptomatic.
    Clinical Sensitivity (Female Patient Status - all specimens)Higher detection with combined specimensCTM + URINE: 90.8% (84.7-96.9%) Asymptomatic; 95.2% (94.0-96.4%) Symptomatic. Total CTM: 89.4% (85.0-93.8%). Total Urine: 86.9% (82.0-91.8%). (Note: The document states "better concordance with culture/DFA positive patients when both swab and urine specimens are tested... resulted in fewer unverified positive test results and higher assay sensitivity")
    Clinical Specificity (Female Patient Status - all specimens)High specificity with combined specimensCTM + URINE: 97.4% (96.4-98.4%) Asymptomatic; 96.4% (95.3-97.5%) Symptomatic. Total CTM: 98.2% (97.6-98.8%). Total Urine: 97.8% (97.1-98.4%).

    2. Sample Size and Data Provenance (Clinical Study)

    • Test Set Sample Size:
      • Total Patients: 4277
      • Total Specimens (initially collected): 8523 (both swab and urine from 4201 patients; urine only from 76 patients)
      • Specimens included in primary analysis (with IC): 8397
      • Specimens included in analysis (without IC): 8478 (after excluding initial equivocal and repeatedly inhibitory results)
    • Data Provenance: Retrospective, collected from six geographically diverse sites. Specific countries are not mentioned, but "US Highway 202 Somerville, New Jersey" for Roche suggests the study was likely conducted in the United States.

    3. Number of Experts and Qualifications for Ground Truth

    • The document does not specify the "number of experts" or their "qualifications" involved in establishing the ground truth.
    • The ground truth method used was standard cell culture with immunofluorescent staining for C. trachomatis for positive samples.
    • For swab specimens that were culture negative by the COBAS test, DFA (Direct Fluorescent Antibody) testing was performed for the presence of C. trachomatis. These are established laboratory methods.

    4. Adjudication Method for the Test Set

    • The document describes a hierarchical adjudication method for establishing the ground truth, rather than a traditional consensus by independent experts on the device's results.
    • Primary Ground Truth: C. trachomatis cell culture (with cyclohexamide treated McCoy cells stained with fluorescein-labeled monoclonal antibody).
    • Secondary Ground Truth (for culture-negative COBAS-positive swabs): DFA (Direct Fluorescent Antibody) for C. trachomatis.
    • Tertiary Check (for information only): Alternate PCR testing (MOMP gene target) was performed on COBAS AMPLICOR positive, culture/DFA negative specimens, but these results were explicitly not used to calculate the clinical performance characteristics. This indicates a strict reliance on the established culture/DFA as the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done in the context of human readers directly interacting with or improving with AI assistance. This document describes the performance of a standalone in vitro diagnostic (PCR-based) test. The "readers" here are the laboratory technicians conducting and interpreting the instrument's output.

    6. Standalone Performance

    • Yes, a standalone study was done. The entire clinical performance section (Tables 3-7 and associated text) describes the performance of the COBAS AMPLICOR CT/NG Test as a diagnostic device without human-in-the-loop assistance in its interpretation beyond standard laboratory procedures (e.g., reading optical density and confirming results). It directly compares the device's output to the established gold standard.

    7. Type of Ground Truth Used

    • The primary ground truth used was a combination of expert-performed laboratory methods:
      • Cell Culture (with immunofluorescent staining) for C. trachomatis.
      • Direct Fluorescent Antibody (DFA) for C. trachomatis (used for culture-negative COBAS-positive samples).
    • The MOMP PCR assay was an investigational check for false positives, but its results were not incorporated into the official performance metrics, highlighting adherence to established methods for ground truth.

    8. Sample Size for the Training Set

    • The document does not provide details on a separate "training set" as would be typical for machine learning algorithms. This device is an in vitro diagnostic (IVD) test, and its development typically involves internal validation and optimization against characterized samples, followed by clinical validation. The data presented is primarily for the clinical validation.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no explicit "training set" discussed in the context of this 510(k) summary for an IVD, the method for establishing its ground truth is not detailed. However, for a device like this, the development process would have involved using characterized C. trachomatis strains and clinical samples with known infection status (determined by culture, etc.) to optimize the assay's primers, probes, and reaction conditions.